2-10
764333-675
Use right or left arrows (< or >) to select pressure units.
Press
SAVE VALUES
touch pad to continue. Display
shows:
ENABLE PC?
YES NO
(PC) selection is used for PC option. This requires a
separate kit. If PC is to be enabled, select
YES
. If not
enabled, select
NO
. If
NO
is selected, display advances
to
LANGUAGES
. Select
YES
if connected to PC. Press
SAVE VALUES
touch pad to continue. Display shows:
PC NODE ADDRESS=
XX
This is used for PC applications. Follow PC kit
instructions to identify node address. Enter correct
address by using number keypad. Press
SAVE
VALUES
touch pad to continue.
Display shows:
LANGUAGE
ENGLISH
Use left or right arrows (< or >) to select desired
language. Press
SAVE VALUES
touch pad. Display
shows:
* UPDATING *
* MEMORY *
The printer prints: * Change Value X:XX:XXA or P. See
Figure 2-5. The printer prints values selected in normal
Operating Mode.
2.1.6 Calibration
The Calibration section of the Service Mode has four
sub-sections:
• CALIBRATION routine – for adjusting pressure,
temperature, concentration and UV system.
• RESTORE routine – to review or enter calibration
data.
• VIEW routines – to review real time in each of the
analog channels.
• PRINT routine – for printouts of current calibration
data.
• Calibration
NOTE: During the Calibration routine, when the control
displays: STABILIZING, the control is waiting for the
temperature or pressure to be at a stable point to allow
for setting the correct value. If the control remains in
STABILIZING and does not proceed, it would indicate
that there is a fault that must be repaired before
calibration can be accomplished successfully.
When the display shows:
MFG TESTS
VALUES CALIB
Select
CALIB
, then press
SAVE VALUES
touch pad.
Display shows:
CALIB RESTORE
VIEW PRINT
Select
CALIB
, then press
SAVE VALUES
touch pad.
Display shows:
PRESS TEMP
CONC MORE
S E T U P
D-MM-YY H:MM:SSP
ACCESS CODE: NO
OPERATOR ID: NO
PATIENT ID: NO
DEVICE ID: NO
PROCEDURE ID: NO
PHYSICIAN ID: NO
INFORMATIVE OPTION: NO
DUPLICATE PRINT: NO
COMPLETE TONE: YES
MAXPURE FILTER
FILTER START DATE: D-MM-YY
CYCLE COUNT XXXX
-------------------------
Figure 2-5. Setup
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......