5-17
764333-675
5.4.6
Printer Prints:
TC3 FAILURE
Occurs if the drain block thermocouple
reading is greater than 75°C or less than
0°C in DIAGNOSTIC Cycle.
1. Check TC3 for loose
connections or broken
wires.
2. Check analog cable
connections. (Refer to
Figure 5-5.)
3. Re-enter known good
calibration values.
4. Defective TC3. Replace.
5. Defective interface board.
Replace board.
5.4.7
Printer Prints:
CHM-PRES
NOT
CALIBRATED
Indicates PT1 has not been calibrated or
calibration data lost and default values
are being used.
1. Calibrate the PT1.
2. Re-enter known good
calibration values.
3. Defective battery-backed
RAM. Replace RAM.
5.4.8
Printer Prints:
INLET NOT
CALIBRATED
Indicates TC1 (inlet thermocouple) has
not been calibrated or calibration data
lost and default values are being used.
1. Re-enter known good
calibration values
2. Calibrate TC1.
3. Defective battery-backed
RAM. Replace RAM.
5.4.9
Printer Prints:
OUTLET NOT
CALIBRATED
Indicates TC2 (outlet thermocouple) has
not been calibrated or calibration data
lost and default values are being used.
1. Re-enter known good
calibration values.
2. Calibrate TC2.
3. Defective battery-backed
RAM. Replace RAM.
5.4.10 Printer Prints:
DRN-BLK NOT
CALIBRATED
Indicates TC3 (drain block
thermocouple) has not been calibrated
or calibration data lost and default
values are being used.
1. Re-enter known good
calibration values.
2. Calibrate TC3.
3. Defective battery-backed
RAM. Replace RAM.
5.4.11 Printer Prints:
CONC. NOT
CALIBRATED
Indicates concentration monitor has not
been calibrated or calibration data lost
and default values are being used.
1. Calibrate the
concentration monitor.
2. Re-enter known good
calibration values.
3. Defective battery-backed
RAM. Replace RAM.
5.4.12 Printer Prints:
FLASH VALUES
RESTORED
Flash Memory Parameters Restored.
Occurs on power-up if battery-backed
RAM values are lost. RAM is initialized.
Flash ROM values are restored.
1. Battery Backed RAM
failure. Replace RAM.
2. Main control board failure.
Replace board.
5.4 CONTROL FAULTS (CONTINUED)
Fault
Description
Service Instructions
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......