6-3
764333-675
6.4.9
CK8 Test
• This test can be used to determine the setpoint of
CK8.
• This test requires manipulation of components in
the VALVES section of the Service Mode.
1. Turn ON s8 (SOL8).
2. Turn ON s1 (SOL1). Compressor runs to
compressor control switch(LS5) setting.
3. When compressor has stabilized, turn OFF s1
(SOL1).
4. Turn OFF s8 (SOL8).
5. Press STATUS touch pad, and monitor system
pressure. The reading that it stabilizes at is the CK8
setting. Acceptable range is greater than 3.5 psi or
less than 11.7 psi.
6.4.10 Output Test
This test is used to operate the board outputs.
Enter the Service Mode, select TESTS, then select
VALVES and the following outputs can be operated:
tk – Paper take up
s8 – Air-water solenoid valve (SOL8)
s5 – Dump valve (SOL5)
s3 – Circulation pinch valve solenoid (SOL3)
s4 – Drain pinch valve solenoid (SOL4)
s7 – Inlet fill valve (SOL7)
s2 – Seal release solenoid (SOL2)
s1 – Compressor release solenoid (SOL1) and enable
compressor (COMP1)
h – Heater element (HTR1)
p1 – Circulation pump
p2 – High pressure pump (HP Pump)
6.4.11 Touch Pad Test
In Service Mode, select
TESTS
, then
CONSOLE
, then
KEYPAD
. This brings up keypad test which indicates
a number on display when corresponding pad is
pressed. If display does not change when pad is
pressed, it indicates a faulty touch pad.
• Console
CONSOLE allows checking of Keypad, Dip Switch
settings, and Display.
Select
CONSOLE
, then press
SAVE VALUES
touch
pad. Display shows:
KEYPAD DIPSW
DISPLAY
• Keypad
Select KEYPAD, then press SAVE VALUES touch pad.
Display shows:
* KEYPAD TEST *
KEY =
Refer to the illustration below. Press a key on the
keypad. If the communication link is OK, the
corresponding number on the keypad displays.
NOTE: The CANCEL touch pad does not display a
number. The test for the CANCEL touch pad is that it
exits this routine when pressed.
Keypad Test Chart
0 1 2
3 4 5
6 7 8
9
15 10
16 17 18 1
9
20
21
XXXXXXXXX
S
TART
23
CANCEL
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......