7-27
764333-675
7.27 RESET UV LAMP ELAPSE-TIME
METER
The elapsed-time meter measures the number of
months that the lamp has been operating. The lamp
must be replaced after it has been operating for 12
months. After 11 months of operation, the time meter
indicator turns yellow. At 12 months of operation the
meter sounds an audible alarm and the indicator turns
red.
To reset the elapsed-time meter follow these steps.
1. Disconnect the power supply power cord from the
wall receptacle and leave it unplugged for 10
seconds.
2. Press and hold the lamp reset pushbutton.
3. Plug the power cord into the wall receptacle while
continuing to press and hold the lamp reset
pushbutton. The indicator light flashes green for
three seconds then indicator light flashes red. As
soon as indicator light flashes red, immediately
release the lamp reset pushbutton.
7.28 CONVERT FAN FROM VERTICAL
TO HORIZONTAL MOUNTING
CONFIGURATION
The UV System can be mounted either in a vertical or
horizontal orientation. The UV water treatment system
is shipped as a vertically configured system and the
fan is mounted to the S/S chamber as shown in Figure
7-40. If the UV system will be mounted horizontally,
the fan assembly needs relocated as described in the
following steps.
1. Remove fan assembly from the UV chamber by
unthreading the 4" hose clamp and discard.
2. With a small flat blade screwdriver, lift the tabs on
both sides of the housing and slide the heat sink
from the fan cover as shown in Figure 7-41. Discard
heat sink.
Figure 7-40. Convert Fan from Vertical to
Horizontal Mounting
Vertical Mounted Fan Position
Horizontal Mounted Fan Position
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......