2-9
764333-675
VALVES INPUT
CONSOLE MORE
Select
MORE
, then press
SAVE VALUES
touch pad.
Display shows:
HPP TEST DIAGVAL
LISTS HISTORY
• Lists
Select
LISTS
to delete Operator ID and/or Physician
ID list(s). When display shows: Select
LISTS
, then
press
SAVE VALUES
touch pad. Display shows:
PRINT LISTS
CLEAR LISTS
To print a list, select
PRINT LISTS
option, then press
SAVE VALUES
touch pad. Display shows:
PRINT OPER LIST
PRINT PHYS LIST
Select the appropriate response, then press
SAVE
VALUES
touch pad. The list is printed.
To clear a list, select the
CLEAR LISTS
option, then
press the
SAVE VALUES
touch pad. Display shows:
CLEAR OPER LIST
CLEAR PHYS LIST
Select desired list to clear, then press
SAVE VALUES
touch pad.
• History
Display shows:
PRINT ERRORS
PRINT CYCLE DATA
Select
PRINT ERRORS
to print a listing of non-cycle
related errors. These are non-critical errors that do not
abort a cycle, and would only be displayed in the
Service mode. Select
PRINT CYCLE DATA
to print
last five cycle tapes in an Enhanced Operator Mode
format. Press CANCEL touch pad to continue.
2.1.5 Values
Use cursor to select
VALUES
. Press
SAVE VALUES
touch pad to enter change values routine. These values
are not found in operator Change Value Procedures.
This section includes: Cycle Count, Run Time,
Temperature Units, Pressure Units, Time Set. The Main
Menu display shows:
MFG TESTS
VALUES CALIB
Select
VALUES
, then press
SAVE VALUES
touch pad.
Display shows:
CYCLE-COUNT=
XXXXXXXX
When installing a new Control Board, cycle tapes
should be checked and current cycle count entered into
control. Press
SAVE VALUES
, touch pad to continue.
Display shows:
RUN-TIME=
XXXXXXXX hrs
This is an elapsed time from control startup. If a new
control board is being installed, start time at 0 hours.
Press
SAVE VALUES
touch pad to continue. Display
shows:
Enter Serial Number of unit, using keypad and right
or left arrows (< or >). Press
SAVE VALUES
touch pad
to continue. Display shows:
PRESSURE UNITS
PSI BAR
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......