1-3
764333-675
WARNING – LIQUID CHEMICAL STERILANT VERIFICATION HAZARD:
Failure to thoroughly clean devices prior to processing may result in an ineffective liquid chemical
sterilization process.
Use the Quick Connect specifically designed for the device being processed.
Do not use S40™ Sterilant Concentrate after the expiration date indicated on the label.
Do not substitute any other product for S40 Sterilant Concentrate.
Use of a blocked or damaged aspirator may result in ineffective liquid chemical sterilization.
Always verify that the sterilant container is empty at the end of the cycle.
Devices are not liquid chemically sterilized and/or adequately rinsed when a liquid chemical sterilant
processing cycle is cancelled.
Do not adjust inlet water pressure regulator from factory-set value of 50 psig.
Do not adjust Ultra Violet (UV) system hose lengths.
!
!
!
!
!
!
!
!
!
CAUTION – POSSIBLE EQUIPMENT DAMAGE HAZARD:
Do not energize “h”, control screen designation for heater element (HTR1), unless chamber is full of
water; element failure will occur.
Do not energize “s2”, control screen designation for seal release valve (SOL2), with water in the
chamber; spill will occur.
Do not force lid to close.
After a cycle is cancelled, if fluid remains in the chamber, call STERIS.
When replacing the control board, use care when picking up the sheet metal/board assembly, because
the display and printer cables are still connected under the board.
In the Service Mode, do not energize Heater unless chamber is full of water; element failure will occur.
In the Service Mode, do not energize Seal Relief Valve/Compressor with water in the chamber; spill
will occur.
To avoid jamming paper in printer, do not force paper into printer mechanism and do not attempt to
feed uneven, tattered, wrinkled paper.
!
!
!
!
!
!
!
!
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......