5-14
764333-675
5.2.15 Drain Check Fault
Description
– At end of Diagnostic Cycle, pressure in
housing of the MaxPure Filter is vented through CK8
to check cracking pressure. Pressure must be as follows
or alarm occurs:
3.51psi < pressure < 11.7psi
Action:
1. Faulty CK8 – replace.
2. Drain line restriction – repair.
3. Faulty PT1 – repair. (Refer to Figure 5-1 and 5-2.)
5.2.16 Ambient Temperature Fault
Description
– A sensor on the control board monitors
ambient room temperature and alarm occurs if
temperature measured is > 65°C or <13°C.
Action:
1. Room temperature out of specification – resolve.
2. Faulty control board – replace.
5.2.17 Lid Open Fault
D
escription
– Control monitors lid switch for open
condition during cycle. If open is sensed, alarm occurs.
Action:
1. Faulty LS4 – replace. (Refer to Figure 5-3.)
2. Loose connection at LS4 – repair.
3. Incorrect adjustment of latch flag assembly –
re-adjust.
5.2.18 HP Pump Fault
Description
– Pressure switch LS6 monitors the pump
output under pump ON and OFF conditions. Should
the switch sense incorrect operation, an alarm occurs.
Action:
1. Wrong tray adapter used – refer to operating
instructions (e.g., C1315 adapter to be used on
C1220E and C1140E trays).
2. Faulty LS6 switch – replace.
3. Faulty P2 – test flow rate and output pressure,
replace if out of specifications.
4. Tubing under tray obstructed – replace tray.
5. Incoming water pressure out of specification –
check utility pressure values and discuss with
Customer.
5.2 DIAGNOSTIC CYCLE FAULTS (CONTINUED)
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......