5-56 Source
Commands
SR785 Dynamic Signal Analyzer
Noise Source Commands
NAMP (?) {x} <mV, [V}, dBVpk>
The NAMP command sets (queries) the Noise Amplitude. The parameter x is the
amplitude in the specified units. The query returns two numbers of the form y,i
where y is a real number and i is an index indicating the units.
This command is valid only when the Source Type is Noise.
NTYP (?) {i}
The NTYP command sets (queries) the Noise Type. The parameter i selects
Bandlimited White (0), White (1) or Pink (2).
This command is valid only when the Source Type is Noise.
NBUR (?) {x}
The NBUR command sets (queries) the Noise Burst Percentage. The parameter x
is a percentage from 1 to 100.
This command is valid only when the Source Type is Noise and the
Measurement Group is FFT or Time/Histogram.
NPER (?) {x} <ks, [s], ms>
The NPER command sets (queries) the Noise Burst Source Period. The
parameter x is a time from 4 ms to 1 ks.
This command is valid only when the Source Type is Noise and the
Measurement Group is Octave or Order.
CSRC (?) {i}
The CSRC command sets (queries) the Source Display. The parameter i selects
DisplayA (0) or DisplayB (1)
This command is valid only when the Source Type is Chirp or Noise and the
Measurement Group is FFT.