4-160 Exceedance Statistics Menu
SR785 Dynamic Signal Analyzer
Exceedance Pct
Set the Exceedance Centile (L
n
).
L
n
is the amplitude at each bin which is exceeded by n% of the records in the waterfall.
L
1
is a large amplitude exceeded only 1% of the time. L
99
is a small amplitude exceeded
by 99% of the measurements.
The range of records within the waterfall buffer which are analyzed is specified by the
Start and Stop index.
Both displays use the same Exceedance Centile.
Command: EPCT (?) {i}
Calculate Excd
Start the exceedance centile calculation for the active display. Choose a Trace to store
the result and press [Enter] to proceed.
The display must be paused or done with a 1-shot waterfall. Waterfall display is not
required to be on. The save option must be set to Active Measurement Only.
The result is stored in a data trace and has the same measurement type as the waterfall
measurements. To view the result, recall the trace to a display or reference graph.
Command: EXCE d, i