4-24 Time/Histogram Frequency Menu
SR785 Dynamic Signal Analyzer
Base Frequency
This softkey sets the allowed values of the sampling time. If set to 100 kHz the
fundamental sampling rate of the instrument is 256 kHz and the allowed sampling times
will be multiples of the reciprocal of this frequency (3.91
µ
s). If set to 102.4 kHz the the
fundamental sampling rate of the instrument is 262.144 kHz and the allowed sampling
times will be multiples of the reciprocal of this frequency (3.814
µ
s).
Command: FBAS (?) d {, i}
Repeat
When repeat is set 'off', the SR785 computes a single histogram and stops. If repeat is on,
the analyzer will automatically acquire new histograms continually as each one is
finished. This can be useful for analyzing signals which change slowly with time.
Command: HRPT (?) d {, i}\$