2-64 User Math Functions
SR785 Dynamic Signal Analyzer
IFFT( ) is the inverse complex FFT operation. This operation switches the X axis type.
d/dx( ) is the derivative operator. The derivative is performed with respect to the X-axis
bin number, not the x-axis values. For example, to convert d/dx( ) to d/df( ) for a linear
frequency display, divide by the frequency spacing of the bins. In the case where the X-
axis is logarithmic, the function needs to correct for d(logx)/dx. The d/dx( ) operator
requires a smoothing aperture or window.
GrpDly( ) is the group delay operator. Group delay is simply d
θ
/d
ω
. Use [Display
Options]<d/dx Window> to set the aperture. This operator yields valid results for any
complex operand which has a frequency x-axis.
AWt( ) is the A-Weighting operator. This filter is a software version of the input A-
Weight filter. BWt( ) and CWt( ) are the software B-Weighting and C-Weighting
operators. The AWt( ), BWt( ), and CWt( ) operators should only be used on operands
with a frequency x-axis. All three filters conform to ANSI Standard S1.4-1983.
User Function Limits
User Functions and Constants are defined in the [User Math] menu. Only the 5 functions
within the current Measurement Group are available for editing in this menu.
Editing a function which is currently being displayed is allowed. In this case, Traces
which do not contain compatible measurements are not allowed as operands.
User Functions have a maximum length of 31 operators plus operands. User Functions
cannot use another User Function as an operand.