3-28 Status
Indicators
SR785 Dynamic Signal Analyzer
Control Keys
[Start/Reset]
If the unit is already making measurements, [Start/Reset] resets any averages and starts
the measurement over. If the measurement is paused, [Start/Reset] starts the
measurement over again. The Run/Pause indicator shows ‘Run’ when measurements are
in progress.
[Start/Reset] also puts into effect any measurement related changes which may have been
made since the last measurement was completed. For instance, if a set of averages is
completed and the frequency span is changed, the new span does not take effect until the
next time [Start/Reset] is pressed.
If the measurement is triggered, [Start/Reset] does not supply a trigger. If the trigger is
manual armed, [Start/Reset] does not arm the trigger.
If the displays don’t update, see the Operating Hints in Chapter 2 for help.
A TTL rising edge at the rear-panel 'Start' input is equivalent to pressing the [Start/Reset]
key.
Command: STRT
[Pause/Cont]
If the unit is already making measurements, [Pause/Cont] halts data acquisition. The
measurement is paused where it is and the displays are not updated. The Run/Pause
indicator shows ‘Pause’ when measurements are paused.
If the measurement is paused, [Pause/Cont] resumes the measurement where it was left
off. To start over, press [Start/Reset]. The Run/Pause indicator shows ‘Run’ when
measurements are in progress.
For Exponential averaging, [Pause/Cont] simply resumes data acquisition without
resetting the average. For Linear averaging, [Pause/Cont] will complete the current
average. If the linear average was already complete, [Pause/Cont] will start a new
measurement..
For Swept Sine measurements, [Pause/Cont] simply resumes the sweep where it was
paused.
If measurement related parameters are changed while the unit is paused, the
measurement cannot be resumed. The SR785 will display the message "Can't continue,
Measurement has Changed." In this case it is necessary to press [Start/Reset] to restart
the measurement.
Command: PAUS and CONT
[Start Capture]
[Start Capture] starts storing data in the capture buffer.