5-54 Source
Commands
SR785 Dynamic Signal Analyzer
Sine Source Commands
S1FR (?) {f} <kHz, [Hz], mHz,
µµµµ
Hz>
The S1FR command sets (queries) the Frequency of Sine Tone 1. The parameter
f is a frequency in the specified units. The query command always returns a
value in Hz.
This command is valid only when the Source Type is Sine.
S1AM (?) {x} <mVpk, mVpp, mVrms, [Vpk], Vrms, dBVpk, dBVpp, dBVrms>
The S1AM command sets (queries) the Amplitude of Sine Tone 1. The
parameter x is a real number of the specified units. The query returns two
numbers of the form y,i where y is a real number and i is an index indicating the
units.
This command is valid only when the Source Type is Sine.
SOFF (?) {x} <mV, [V]>
The SOFF command sets (queries) the Offset of the Sine Source. The parameter
x is a real number of the specified units. The query returns two numbers of the
form y,i where y is a real number and i is an index indicating the units.
This command is valid only when the Source Type is Sine.
S2FR (?) {f} <kHz, [Hz], mHz,
µµµµ
Hz>
The S2FR command sets (queries) the Frequency of Sine Tone 2. The parameter
f is a frequency in the specified units. The query command always returns a
value in Hz.
This command is valid only when the Source Type is Sine.
S2AM (?) {x} <mVpk, mVpp, mVrms, [Vpk], Vrms, dBVpk, dBVpp, dBVrms>
The S2AM command sets (queries) the Amplitude of Sine Tone 2. The
parameter x is the amplitude in the specified units. The query returns two
numbers of the form y,i where y is a real number and i is an index indicating the
units.
This command is valid only when the Source Type is Sine.