Limit Test Commands 5-87
SR785 Dynamic Signal Analyzer
Limit Test Commands
LSON (?) d {, i}
The LSON command sets (queries) the Limit Segments Display for display d.
The parameter i selects Hide (0) or Show (1).
LTST (?) d {, i}
The LTST command sets (queries) the Limit Testing for display d. The
parameter i selects Off (0) or On (1).
LALM (?) d {, i}
The LALM command sets (queries) the Limit Fail Alarm for display d. The
parameter i selects Off (0) or On (1).
LFAL ? d
The LFAL ? command queries the result of the most recently completed limit
test of display d. If the test passed, 0 is returned. If the test failed, 1 is returned.
LCLR d
The LCLR command clears the Limit Segments for display d. All defined limit
segments for display d are erased. The limits for each display are cleared
separately (d=2 is not allowed).
This command should always be sent before defining any limit segments. Limit
segments may not be defined or edited after the Measurement or Measurement
Group is changed until LCLR clears the existing segments.
LMAX (?) d {, i}
The LMAX command sets (queries) the Last Limit Segment Number for display
d. The parameter i is a last limit segment number from 0 to 199. The set
command is not valid for d=2 (both displays).
The LMAX ? command queries the last limit segment number.
The LMAX d, i command sets the last limit segment number to i. If i is greater
than the current last limit segment number, new entries are created up to line i.
These entries simply follow the last limit segment. If the value of i is less than
the last line number, an error is returned. To remove entries from the table, use
the LDLT command.
This command also sets Limit Segments to Show.
LSEG (?) d, i {, j, x0, y0, x1, y1}
The LSEG command sets (queries) the Endpoints and Type of limit segment i for
display d. The parameter i selects the limit segment number from 0 to the last
segment. If i exceeds the last segment number (as set by LMAX), an error is
reported.