Window
Commands
5-79
SR785 Dynamic Signal Analyzer
This command is valid only when the Measurement Group is FFT or Order. The
set command requires display d to be Live and the Force Exponential window to
be selected..
FWTC (?) d {, x}
The FWTC command sets (queries) the FFT Expo Window Time Constant for
display d. The parameter x is a percentage of the time record.
This command is valid only when the Measurement Group is FFT or Order. The
set command requires display d to be Live and the Force Exponential window to
be selected..
TRWI i
The TRWI command copies the real part of stored Trace i to the User window.
Trace i must contain stored data.
This command is valid only when the Measurement Group is FFT or Order.
WITR d, i
The WITR copies the window of display d to the real part of Trace i. Trace i will
have a length of 2048 points.
This command is valid only when the Measurement Group is FFT or Order.
WSYM (?) d {, i}
The WSYM sets (queries) User Window Form for display d. The parameter i
selects Non-Symmetric (0) or Symmetric (1).
This command is valid only when the Measurement Group is FFT or Order and
the window of display d is User.
Summary of Contents for SR785
Page 4: ...ii ...
Page 10: ...viii ...
Page 80: ...1 64 Exceedance Statistics ...
Page 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...
Page 536: ...5 136 Example Program SR785 Dynamic Signal Analyzer ...