Swept Sine Measurement 1-41
SR785 Dynamic Signal Analyzer
Swept Sine Measurement
This example investigates the test filter (enclosed with this manual) using Swept Sine measurements.
You will use the SR785 source to provide a sweeping sine source and both inputs to measure the input to
and output from the device under test.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Use a BNC Tee to connect the Source Output to
the filter input and the Ch1 A Input.
Connect the filter output to the Ch2 A Input.
In this instrument, transfer function is defined as
Ch2 response over Ch1 reference. Thus, Ch1
monitors the filter input (source output) and Ch2
measures the response of the device under test.
Press [Input], <Input Conifg>
Select the Input Configuration submenu.
Press <Ch1 Input Range>
Adjust the Ch1 input range.
Press [2], select (dBVpk) with the knob, and
[Enter].
Set the range to 2 dBV.
3. Press [Display Setup]
Select the Display Setup menu.
Press <Measurements Group>
There are six Measurement Groups - FFT,
Correlation, Octave, Order Analysis, Swept Sine,
and Time/Histogram. The Measurement Group
determines which Measurements are available to
the displays.
Select (Swept Sine) with the knob and press
[Enter].
Choose the Swept Sine group. The menus now
configure swept sine measurements only.
4. Press [Freq]
Select the Frequency menu.
Press
<Start>
Adjust the sweep Start Frequency.
Press [9] [0] [0] select (Hz) with the knob and
press [Enter].
Enter 900 Hz.
Press
<Stop>
Adjust the sweep Stop Frequency.