Input
Commands
5-63
SR785 Dynamic Signal Analyzer
EU2M (?) d {,i}
The EU2M command sets (queries) the Ch2 Engineering Units Mode for display
d The set command requires d = 2 (both displays).. The parameter i selects Off
(0) or On (1).
EU2L (?) d {,i}
The EU2L command sets (queries) the Ch2 Engineering Units Label for display
d. The set command requires d = 2 (both displays). The parameter i selects the
unit label.
i Label
i Label
0
m/s
2
8
kg
1 m/s
9 lbs
2
m
10
N
3
in/s
2
11
dyne
4
in/s
12
Pas
5
in
13
bar
6
mil
14
USER
7
g
EU2V (?) d {,x}
The EU2V command sets (queries) the Ch2 Engineering Units per Volt scale for
display d. The set command requires d = 2 (both displays). The parameter x is
real number of EU/Volt (EU’s per Volt).
EU2U (?) d {,s}
The EU2U command sets (queries) the Ch2 User Label for display d. The set
command requires d = 2 (both displays). The string s is the user label.
TD2C (?) d {,i}
The TD2C command sets (queries) the Ch2 Transducer Conversion. for display
d. The set command requires d = 2 (both displays). The parameter i selects the
units to be converted to:
i Label
0
m/s
2
1
m/s
2
m
3
in/s
2
4
in/s
5
in
6
mil
7
g