Sideband Marker Menu 4-63
SR785 Dynamic Signal Analyzer
measurement span which are identified by Sideband Markers contribute to the
calculation.
The Sideband Ratio in dB is 10log( sideband power/(fundamental power+sideband
power) ).
Command: SPWR ? d , i
RMS Sideband Power
This menu box displays the Sideband Power of the active display. The Sideband Power
is the sum of the squared magnitudes of the sidebands. Only those sidebands within the
measurement span which are identified by Sideband Markers contribute to the
calculation. The result is shown in Vrms, or dBVrms depending on the setting of dB
Units. To convert a value in Vrms to power, square the result
Command: SPWR ? d , i
Summary of Contents for SR785
Page 4: ...ii ...
Page 10: ...viii ...
Page 80: ...1 64 Exceedance Statistics ...
Page 158: ...2 78 Curve Fitting and Synthesis SR785 Dynamic Signal Analyzer ...