Sideband Marker Menu 4-61
SR785 Dynamic Signal Analyzer
Sideband Marker Menu
This menu is displayed when the active display Marker Mode is Sideband.
Sideband Separation
Set the Sideband Separation for the active display.
The Sideband Marker identifies the carrier frequency with the Fundamental Marker
(solid vertical line) and the number of sidebands with small triangular Sideband Markers.
The Sideband Separation determines the location of these Sideband Markers relative to
the Fundamental.
Only those sidebands within the measurement span which are identified by the Sideband
Markers contribute to the Sideband Power calculations.
Each display has its own Sideband Separation. This entry field can be linked to both
displays using the [Link] key.
Command: SSEP (?) d {, f}
# Sidebands
Select the Number Of Sidebands for the active display [1..50]. The first upper sideband
(fundseparation) is identified as #1, the first lower sideband (fundamental-
separation) is #-1, etc.
The Sideband Marker identifies the carrier frequency with the Fundamental Marker
(solid vertical line) and the number of sidebands with small triangular Sideband Markers.
The Sideband Separation determines the location of these Sideband Markers relative to
the fundamental.