Frequency
Menu
4-7
SR785 Dynamic Signal Analyzer
Frequency Menus
The Measurement Group determines which Frequency menu is displayed. This menu sets
the frequency parameters for the measurement. A set of parameters is stored for each
group.
FFT Frequency Menu
When the measurement group is FFT or Correlation, this menu sets the frequency span
and location for the active display.
Span
Select the FFT Span of the active display. The frequency Span ranges from the FFT Base
Frequency (102.4 kHz or 100.0 kHz) to 2
-19
times the Base Frequency (195.3 mHz or
191 mHz) in factors of 2. A numerically entered value is rounded to the nearest
allowable Span. If the new frequency Span would extend below 0 Hz or above the Base
Frequency, then the Start and End frequencies will be adjusted.
Changing the Span will change the Linewidth (Span/FFT Resolution) and Acquisition
Time (FFT Resolution/Span). The Linewidth and Acquisition time are other ways to
change the Span.
The two displays can have different FFT Spans if the Analyzer Configuration is set to
Independent Channels. In this configuration, no two channel measurements are allowed
(frequency response, cross-spectrum, etc.), but the entry field can be linked to both
displays using the [Link] key. If Analyzer Configuration is set to Dual Channel the field
is automatically linked to both displays.