3-12 Screen
Display
SR785 Dynamic Signal Analyzer
Figure Chapter 3 -4 Overlay Display Screen
The Overlay display format is shown above. The measurement of the inactive display is
shown in half-intensity. The Single display format is similar but the inactive display is
not shown at all.
Data Graph
Data is normally graphed with signal amplitude on the Y axis and frequency or time on
the X axis. The physical size of the graph remains constant while the vertical and
horizontal scales may be changed. The graph area has a dotted grid for reference. There
are 10 horizontal divisions and either 8, 10, 12 or 15 vertical divisions. The horizontal
span is determined by the measurement span or time. The display normally shows the
entire measurement span. The X axis may be expanded (Zoom) and translated (Pan) to
display less than the full measurement span. This expansion does not change the
measurement span or time, it merely changes the display of the data.
Half Intensity Data Graph
Unsettled measurement data is graphed in half intensity. Once the measurement becomes
settled, the graph returns to full intensity.
FFT and Octave measurements are unsettled when the frequency span or resolution is
changed and when the input signal path is changed (gain, filters, etc.). The measurement
is not settled again until the discontinuity in the input data has propagated through the
digital filters and, for FFT’s, a complete new time record has been acquired.
FFT measurements running with a Time Record Increment less than 100% (overlapped
time records) display unsettled measurements in half intensity.