Exceedance
Statistics
1-63
Press
[Pause/Cont]
Continue the live measurement. Note that the live
measurement data rarely exceeds the reference
graph.
9. Press [Pause/Cont]
Pause the live measurement again.
Press <Exceedance Pct>
Change the Exceedance Percentage.
Press [9] [9] [Enter].
Enter 99%. This level is exceeded 99% of the time.
Press <Calculate Excd>, choose (Trace 2) with
the knob and press [Enter].
The exceedance results are stored in Trace 2.
Press [Alt] [Link], choose (Trace 2) with the
knob and press [Enter].
Trace to Ref is an alternate function. The alternate
key functions are labeled below each key (in this
case, the [Link] key).
Trace to Ref copies the data in a trace to the
reference graph of the active display. The reference
graph is shown in half intensity.
In this case, the reference graph is the level of each
octave bin which is exceeded 99% of the time (L
99
).
This concludes this example. Exceedance is a
common measurement for environmental noise
levels, such as airports or highways. Change the
octave integration time and waterfall storage
interval to optimize the measurement rate. Use a
larger waterfall buffer (up to 2000 measurements
per display) for long monitoring times.