Noise Source Menu 4-75
SR785 Dynamic Signal Analyzer
Noise Source Menu
The Noise Source menu is used to configure the noise source.
Amplitude
Set the peak Amplitude of the noise output. The source output will overshoot this
amplitude by as much as 100% a small percentage of the time. Because of the nature of
noise, the peak amplitude is not perfectly defined.
Command: NAMP (?) {x}
Type
Select the Type of noise output. [White, Pink, Bandlimited White or Pink].
White Noise
White noise provides equal noise density (V/
√
Hz) from 0 to above 102.4 kHz (regardless
of the measurement bandwidth of the displays). The spectrum of white noise appears flat
in an FFT spectrum. The power spectral density (PSD) remains constant at all spans.
Pink Noise
Pink noise rolls off at 3 dB/oct providing equal energy per octave and extends beyond
102 kHz. The spectrum of pink noise appears flat in octave analysis.