1-36 Waterfall
Display
SR785 Dynamic Signal Analyzer
Use the knob to select (Start) and press [Enter].
Choose Start. This means that a new measurement
is started as soon as the previous average is
complete. In this case, a new measurement is made
every 8 ms, regardless of triggering.
If we choose One Lin Avg., then a measurement is
made only when triggered.
Press <Power Bin>
Change the Power Bin.
Use the knob to select (L) and press [Enter].
Choose L (L
eq
) as the sound level bin. L
eq
is a
standard broad band sound measurement. When we
start the next measurement the result will be
displayed as the last bin in the display and labeled
‘L’.
7. Press [Trigger]
Select the Trigger menu.
Press <Trigger Source>
Change the Trigger Source.
Use the knob to select (Manual) and press
[Enter].
Choose Manual trigger. We will start our
measurement with a button press. We could also
use an external trigger. Triggering on the signal
itself requires the use of an external signal source.
Press <Trigd Source Mode>
Change the Triggered Source Mode. This
determines whether the source triggers only once
(on the first trigger after [Start/Reset]) or on every
trigger.
Use the knob to select (One Shot) and press
[Enter].
Choose One Shot to trigger the noise burst only
once at the start of the measurement.
8. Press [Start/Reset]
[Start/Reset] starts the measurement. Since the
measurement is triggered (not Free Run), nothing
happens until the first trigger is received.
Press <Manual Trigger>
<Manual Trigger> supplies the first trigger. The
source outputs a single noise burst (Trigd Source
Mode=One Shot). The display starts a continuous
stream of octave measurements, each linear
averaged for 8 ms and each starting when the
previous average is complete (Linear Avg
Mode=Continuous).