Waterfall
Menu
4-135
SR785 Dynamic Signal Analyzer
Waterfall Menu
The Waterfall menu configures the waterfall displays. Waterfall displays are not
available for Nyquist or Nichols Views or Swept Sine measurements.
Display
Select the Normal or Waterfall Display for the active display. Waterfall Display only
applies to FFT and Octave measurements. Waterfall displays are not available for
Nyquist or Nichols Views or Swept Sine measurements.
Normal Display only shows the results of a single measurement.
Waterfall Display WITHOUT storage is simply a graphical display scrolling. The marker
is constrained to the most recent record (at the back of the waterfall). Older records are
only shown graphically, their data has not been saved and may not be accessed in any
way. In this case, the display updates and scrolls as quickly as possible.
Waterfall display WITH storage shows only those records which are stored in waterfall
memory. While the measurement is running, the display scrolls down and the newest
record is added to the top (back). While running, the display may not update fast enough
to show every stored record in real time. Pause the measurement to redraw the display
with either the newest record at the top (normal) or the oldest record at the top
(reversed). When paused, the display shows every stored record which is visible. The
number of records stored in waterfall memory is displayed in the Vertical Scale Bar
when storage is on.
A waterfall display may not be taken Off-Line.
Moving the Marker
The marker may be moved to a record other than 0 only if storage is on.