4-132 Window Menu
SR785 Dynamic Signal Analyzer
window can limit the performance of the analyzer when looking at signals close together
in frequency and very different in amplitude. The Hanning window is most often used in
noise measurements since it has the lowest noise floor.
Flattop
The Flattop window has the best amplitude accuracy of any window. Its off-bin
amplitude variation is about 0.02 dB. However, the selectivity is worse. Unlike the other
windows, the Flattop window has a very wide pass band and very steep rolloff on either
side. Thus, signals appear wide but do not leak across the whole spectrum. The Flattop
window is the best window to use for accurate amplitude measurements.
BMH
The BMH window has reasonable off-bin amplitude accuracy (about 0.8 dB) and much
lower side-lobes than the Hanning window. Thus, it has very good selectivity and very
little broadening of non-bin frequencies. The BMH window is a good window to use for
measurements requiring a large dynamic range.
Kaiser
The Kaiser window has the lowest side-lobes and least broadening for non-bin
frequencies. This makes this window the best for selectivity. The off-bin amplitude
variation is the same as the BMH (about 0.8 dB). The Kaiser window is the best window
to use for measurements requiring a large dynamic range.
Force-Exponential
The Force window is uniform over the beginning of the time record and zero over the
remainder. The force length is user specified. This window is used to isolate impulsive
signals, such as impact excitations, from noise and other oscillations later in the time
record.
The Exponential window attenuates the time record with a decaying exponential time
constant. This window is often used in impact testing on the response channel to remove
oscillations which last longer than the time record.
Many impact measurements require the Force window for the excitation channel and the
Exponential window for the response channel. Use the <Channel 1 Window> and
<Channel 2 Window> softkeys to select which channel to apply the force and
exponential windows to.
User Defined
The User window is any function that the user provides. The User window is copied
from a stored trace using <Trace to Window>. The trace may contain stored data or may
be loaded from disk or via the computer interface.
Remember, window functions have a great deal of impact on the resulting FFT spectrum.
A poorly designed window can result in significant measurement errors.