2-54 Swept Sine Measurements
SR785 Dynamic Signal Analyzer
channel measures an input which exceeds the Ideal Reference by more than the Upper
Limit or which is less than the Ideal Reference by more than the Lower Limit. In cases
where the reference channel input must be limited, the Ideal Reference plus the Upper
Limit must be less than or equal to this limit.
Whenever Auto Level is used, the measurement MUST be frequency response. The
signal connections must be made for a frequency response measurement. This is because
source level changes are not normalized in the individual measurements. Only the ratio
of Channel 2 to Channel 1 is source level independent.
The Source Ramp Rate is the rate at which the source level changes. If Source Ramping
is Off, source level changes are made instantly. If Source Ramping is On, the source
level changes are made at the ramp rate. The Settle Time starts after the source reaches
the new level. This prevents fast level changes from disturbing the device under test. Of
course, this adds to the sweep time.
The source ramps off whenever there are no measurements being made. This occurs
before the start of a sweep, at the end of a single sweep or while a sweep is paused (by
the user). If Ramping is Off, the source shuts off instantly in these situations.