3 S y s t e m C o n t r o l s
[ 1 ] I N S T R U C T I O N S F O R U S E
3 - 4 1
Color Flow LCD Selections
When Color Flow is active, you can access the related LCD selections by
rotating the
FUNCTION SELECT
control on the LCD panel to highlight the
C
tab on the LCD.
LCD Selection
Description
Settings
|
Window
Adjusts the position or size of the color ROI using
the trackball.
POS (position) or Size
|
2D FOV
Reduces the field of view (FOV) for the 2D image to
equal the width of the color ROI by using the trackball.
Off, FOV, POS
|
PSP/QSP
(Available with the SONOLINE G60 S system; requires the
Cardiac Imaging option for the SONOLINE G50 system)
Selects
PSP
(Parallel Signal Processing),
QSP
(Quad
Signal Processing), or
Off
. The Color PSP is independent
of the 2D-mode PSP.
The PSP setting on the SONOLINE G50 system requires
the use of a phased array transducer.
The QSP setting is only available for phased array
transducers.
Available for the SONOLINE G60 S
system only.
Off, PSP
|
Persistence
Causes the signals obtained from blood flow to remain
in the color ROI for longer periods of time when the
setting is increased.
Off, 1, 2, 3, 4, 5, 6, or 7
|
Flow Stage
Allows selection of system settings that are closely
optimized for hemodynamic flow conditions.
H – high velocity and low resistance
HM – medium high velocity
M – medium velocity and high resistance
ML – medium low velocity
L – low velocity and low resistance
H, HM, M, ML or L
|
Density
Selects the density of acoustic lines passing through the
color ROI. Increasing the line density increases
resolution and decreases frame rate.
1, 2, 3, 4, or 5
|
Peak
Designates the period of time that peak color velocities
are accumulated for display.
Off, 1sec, 2sec, or 3sec
|
Vel Tag
Specifies a velocity or a range of velocities to display in a
bright green color. The color designation can help to
detect clinically significant flow stages.
Off, Rng, Point
|
Color Map
Selects the range of colors that are applied to the
velocity range in mapping blood flow.
VV1, VV2, VV3, VV4, A, B,
C, or D
Summary of Contents for SONOLINE G50
Page 9: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Page 19: ...1 Introduction 1 12 1 I N S T R U C T I O N S F O R U S E ...
Page 57: ...2 Safety and Care 2 38 1 I N S T R U C T I O N S F O R U S E ...
Page 143: ...3 System Controls 3 86 1 I N S T R U C T I O N S F O R U S E ...
Page 145: ...4 System Setup 4 2 1 I N S T R U C T I O N S F O R U S E ...
Page 183: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Page 229: ...6 Technical Description 6 28 1 I N S T R U C T I O N S F O R U S E ...