2 S a f e t y a n d C a r e
[ 1 ] I N S T R U C T I O N S F O R U S E
2 - 1 7
Ultrasound System
Care
It is the responsibility of the user to verify that the ultrasound system is safe
for diagnostic operation on a daily basis. Each day, prior to using the system,
perform each of the steps in the Daily Checklist.
All exterior parts of the system, including the control panel, keyboard,
transducers, and biopsy devices, should be cleaned and/or disinfected as
necessary or between uses. Clean each component to remove any surface
particles. Disinfect components to kill vegetative organisms and viruses.
The air filter on the ultrasound system must be cleaned regularly to maintain
proper system cooling. Remove and check the air filter weekly, and clean
as needed.
Daily Checklist
Perform the following each day before using the ultrasound system:
Visually inspect all transducers. Do not use a transducer which has a
cracked, punctured, or discolored casing or a frayed cable.
Discoloration Exception:
The use of Cidex OPA and Gigasept FF disinfectants may
cause discoloration of transducer housings. You can continue to use a transducer if it
is discolored due to the use of these specific disinfectants only.
Visually inspect all power cords. Do not turn on the power if a cord is
frayed or split, or shows signs of wear.
If your system's power cord is frayed or split, or shows signs of wear,
contact your Siemens service representative for power cord
replacement.
Verify that the trackball, DGC slide controls, and other controls on the
control panel and LCD are clean and free from gel or other
contaminants.
Once the system is powered on:
Visually check the on-screen displays and lighting.
Verify that the monitor displays the current date and time.
Verify that the transducer identification and indicated frequency are
correct for the active transducer.
[1] Instructions for Use
Air Filters
2-23
Approved List
of Disinfectants
2-30
Summary of Contents for SONOLINE G50
Page 9: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Page 19: ...1 Introduction 1 12 1 I N S T R U C T I O N S F O R U S E ...
Page 57: ...2 Safety and Care 2 38 1 I N S T R U C T I O N S F O R U S E ...
Page 143: ...3 System Controls 3 86 1 I N S T R U C T I O N S F O R U S E ...
Page 145: ...4 System Setup 4 2 1 I N S T R U C T I O N S F O R U S E ...
Page 183: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Page 229: ...6 Technical Description 6 28 1 I N S T R U C T I O N S F O R U S E ...