3 S y s t e m C o n t r o l s
3 - 4 4
[ 1 ] I N S T R U C T I O N S F O R U S E
LCD Selection
Description
Settings
|
Motion Cut
Increases or decreases the threshold of the high pass filter.
Higher settings provide more filtering and reduction of motion
artifacts. Lower settings provide better sensitivity to lower
velocities and small vessel detection, with minimal reduction
of motion artifacts.
1, 2, 3, or 4
|
Noise Cut
Increases or decreases the value of the spatial filter, which
reduces the random noise level, by looking at adjacent pixels
to determine if a value is real or the result of noise.
Off, 1, 2, 3, 4, 5, 6, 7,
8, or 9
|
Temporal
Averages velocity values for a number of transmit pulses
per acoustic line to optimize color flow signal-to-noise ratio.
Higher values of this setting increase sensitivity, but decrease
frame rate.
1, 2, or 3
|
C Priority
Optimizes color over 2D-mode.
0 (2D-mode), 1, 2, 3,
4, 5, 6, 7, 8, 9, 10, 11,
or 12 (color)
CINE LCD Selections
When CINE is active, you can access the related LCD selections by rotating the
FUNCTION SELECT
control on the LCD panel to highlight the
CIN
tab on the LCD.
LCD Selection
Description
|
Frame Review
Displays individual frames in a forward or reverse direction, in response to the trackball.
Frame numbers are displayed.
For motion modes (M or Doppler), one frame of time motion data is displayed at a time.
No frame numbers are displayed.
|
Motion Review
Displays CINE data in a continuous, forward direction. Reverse motion is not available.
|
Edit Start
|
Edit End
Defines new beginning and ending points of a loop of CINE data.
|
Edit Reset
Resets the beginning and ending points to their originally acquired positions.
|
Align
Synchronizes the CINE display of the dual images during Dual mode.
|
Rate
Changes the speed of CINE review while in Motion Review.
Note:
The signal of a Doppler spectrum is not audible when the review speed has been
adjusted with the
|
Rate
LCD selection.
|
Frame Recall
Activates Frame Recall memory and allows retrieval of individual frames. To enable the
Frame Recall function, use the system presets to select CINE storage as the
destination associated with the
DIGITAL STORE
key.
Summary of Contents for SONOLINE G50
Page 9: ...1 Introduction 1 2 1 I N S T R U C T I O N S F O R U S E ...
Page 19: ...1 Introduction 1 12 1 I N S T R U C T I O N S F O R U S E ...
Page 57: ...2 Safety and Care 2 38 1 I N S T R U C T I O N S F O R U S E ...
Page 143: ...3 System Controls 3 86 1 I N S T R U C T I O N S F O R U S E ...
Page 145: ...4 System Setup 4 2 1 I N S T R U C T I O N S F O R U S E ...
Page 183: ...5 Beginning an Examination 5 2 1 I N S T R U C T I O N S F O R U S E ...
Page 229: ...6 Technical Description 6 28 1 I N S T R U C T I O N S F O R U S E ...