Step 2: Automatic Assignment of Accura 2350 Module IDs and Update of IDs
Accura 2300S automatically assigns IDs to all connected Accura 2350 modules as shown in the
following steps and updates the module IDs. Refer to
「
Chapter 4 Accura 2300S Operation/Setup >
Setup Mode > Accura 2350 Module ID Setup
」
for details about how to automatically assign IDs to
connected modules.
Step 2-1 Automatic Assignment of Standby to Accura 2350 Modules
If the standby IDs
1
are already assigned to all connected Accura 2350 modules, move to step 2-2.
Automatic Assignment of Module IDs
①
Long press
2
the SETUP button.
→
Go to the setup mode.
②
Press the Left(<) or Right(>) arrow button to move to the
「
IduP
」
(module ID
update) screen in the
「
MODULE
」
column.
①
Press the Down(
v
) button to move to the
「
IdSt
」
(automatic ID assignment)
screen.
④
Long press the ENTER button.
→
The text
「
RJ-1 Inc2(RJ12-1/Increment 2)
」
blinks
on the screen.
⑤
Select
「
Stby(Standby)
」
by using the Up(
^
) or Down(
v
) arrow button.
⑥
Long press the ENTER button.
→
The text
「
Stby(Standby)
」
stops blinking.
→
Standby IDs are automatically assigned to all connected Accura 2350 modules.
⑦
Long press the ESC button.
→
Exit the setup mode and move to the display
mode.
1. The standby ID 255 is assigned to all Accura 2350 modules as a factory default value.
2. Press the button for one second. A beep sound occurs in one second.