Fig 2.29 Digital Input
Caution
Since 5 V is output from the dry contact terminal internally, if the external power is applied, it can
damage the meter.
1 Digital Output Channel
Channel 1
Port name
DO
Ground
DO
G
Item
Description
Connector type
Terminal block
1
Wire specifications
1.5
–
3.5 mm
2
(16
–
12 AWG), copper or aluminum
Output contact
1 channel, Form A relay
Output type
2
Latch, Uncountable/Countable single pulse
3
, Periodic pulse
3
Turn-on time
Maximum 6 ms
Turn-off time
Maximum 3 ms
Isolation
AC 2,000 V for 1 minute
Maximum rating
AC 250 V 5 A, DC 110 V 0.3 A, DC 30 V 5 A
1. The maximum screw tightening torque of the barrier-type terminal block for digital input/output ports is 0.9 Nm
(9 kgf-cm, 8 lbf-in).
2. The operations of different types of digital outputs are as follows.
Output Type
Command
Operation
4
Latch
1
The contact remains closed.
0
The contact remains open.
Single pulse(uncountable) 1
A single pulse occurs when the contact is closed. The
command issued after the occurrence of a single pulse is
cleared automatically. Additional commands issued within
the pulse period are ignored.
0
Not available
Single pulse(countable)
1
Single pulse occurs when the contact is closed. The
command issued after the occurrence of a single pulse is
cleared automatically. Additional commands issued within
the pulse period are counted and the counted number of
pulses occur after the end of the pulse period.