Chapter 4 Accura 2300S Operation/Setup
Accura 2300S/2350-1P3FSC User Guide
Page 82
ⓒ
2015 Rootech Inc. All Rights Reserved
Button Operation
Button
Description
[Display Mode]
Press
1
Select the display screen using the up, down, left and right
arrow buttons.
[Module ID Selection]
Long press
2
Make the module ID changeable(the ID blinks at 0.4-
second intervals).
Press
Select the module ID to display
3
.
Long press or
press
Set the module ID and move to the display mode(the
screen stops blinking).
1. When the button is pressed, a beeping sound occurs.
2. Press for more than 1 second. After one second, a beeping sound occurs.
3. If the module ID is changed, the measurement data of the selected Accura 2350 module are displayed on the LCD
screen.
Note
The module ID blinks in the following cases:
①
Blinks at 1.2 second intervals when Accura 2300S cannot properly communicate with Accura
2350 due to
- the disconnection of the communication cable(6C telephone line)
- the change of the module ID
- duplicate IDs
②
Blinks at 0.4 second intervals when the module ID can be changed.
Summary of Contents for ACCURA 2300S