R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Performing Spectrum Measurements
Operating Manual 1309.6946.12 - 09
82
4.1.2.5 Measuring the Maximum Channel Power
If signal levels fluctuate significantly, you can define the maximum channel power with
the Max Hold function.
► Press the MEAS key.
► Press the "Power Display" softkey.
► Select the "Max Hold" menu item.
The power display switches from "Power" to "Max Power".
If you want to deactivate the Max Hold function and return to the normal power display,
activate Clear/Write.
► To deactivate the Max Hold function, press the "Power Display" softkey.
► Select the "Clear/Write" menu item.
The power display returns to "Power" display.
4.1.2.6 Unit for Power Display
The R&S ZVH can apply different units to the power output. The default unit is dBm.
► Press the MEAS key.
► Press the "Power Unit" softkey.
► Select the unit you want from the submenu.
The R&S ZVH displays the power level in the selected unit.