R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Configuring Spectrum Measurements
Operating Manual 1309.6946.12 - 09
124
4.2.1.4 Defining a Start and Stop Frequency
Defining a start an stop frequency is best suited for example for measurements on
harmonics or signals whose exact frequency is unknown.
► Press the FREQ/DIST key.
► Press the "Start Freq" softkey.
The R&S ZVH opens an input field to define the start frequency.
► Enter the start frequency you need.
► Set a stop frequency with the "Stop Freq" softkey.
The R&S ZVH adjusts the horizontal axis according to your input, beginning with
the start frequency and ending with the stop frequency.
If you have entered a stop frequency that is outside the maximum frequency range,
the R&S ZVH sets the stop frequency to the possible maximum.
The label of the horizontal axis changes from "Center" and "Span" to "Start" and
"Stop".
4.2.1.5 Setting the Span
The span is the frequency range around the center frequency that a spectrum analyzer
displays on the screen. The span you should select depends on the signal and the
measurement that you are performing. A rule of thumb is that it should be at least twice
the bandwidth occupied by the signal.
The available span for frequency domain measurements depends on the instrument
model.
● R&S ZVH4: 10 Hz to 3.6 GHz
● R&S ZVH8: 10 Hz to 8 GHz
If you set a span of 0 Hz (zero span), the R&S ZVH performs measurements in the
time domain.
► Press the FORMAT/SPAN key.
When you press the FORMAT/SPAN key, the R&S ZVH automatically opens an
input field to define the span. If the input field is inactive, you can open it with the
"Manual Span" softkey.
► Enter the span you need.
The R&S ZVH adjusts the span of the horizontal axis.