R&S ZVH
Cable and Antenna Test Mode
Configuring Cable and Antenna Tests
Operating Manual 1309.6946.12 - 09
56
2.2.4 Configuring the Tracking Generator
If your R&S ZVH features a tracking generator, you can define several characteristics
of the generated signal.
► Press the SCALE/AMPT key.
► Press the "TG Power" softkey.
► Enter the desired output level of the signal in dBm.
Tracking generator output level
Using an output level of 0 dBm (or a receiver attenuation of 0 dB) for S12 or S21
measurements can cause an overload at the RF input. It is therefore recommended to
use an output level of -10 dBm (default) or lower for these measurements.
This especially applies to measurements on amplifiers.