![Rohde & Schwarz ZVH-K1 Operating Manual Download Page 199](http://html.mh-extra.com/html/rohde-and-schwarz/zvh-k1/zvh-k1_operating-manual_1477901199.webp)
R&S ZVH
Network Analyzer Mode (R&S ZVH-K42)
Performing Vector Measurements
Operating Manual 1309.6946.12 - 09
184
The R&S ZVH shows the results of
the vector transmission
measurement and therefore the
filter characteristics.
You can change the measurement
configuration (e.g. the sweep time
or detector) or format (e.g. view
phase characteristics) without
affecting the accuracy of the
measurement.
For more information see "
Selecting the Measurement Format
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.
Depending on the calibration method, you can also select other result displays
(e.g. reflection measurements) without having to recalibrate the R&S ZVH.