R&S ZVH
Cable and Antenna Test Mode
Configuring Cable and Antenna Tests
Operating Manual 1309.6946.12 - 09
52
Setting the displayed frequency range
After you have selected the span, you can set a particular frequency range whose
results the R&S ZVH displays.
In the default configuration, the R&S ZVH adjusts the DTF start and DTF stop
frequency according to the span and the center frequency. The distance from center
frequency to the start and stop frequency is the same. Alternatively, you can directly
set a DTF start and DTF stop frequency.
► Press the FREQ/DIST key.
► Press the "DTF Center Freq" softkey.
► Enter the frequency that you'd like to be in the center of the horizontal axis.
The R&S ZVH adjusts the frequency range according to span and center
frequency.
► Press the FREQ/DIST key.
► Press the "DTF Start Freq" or the "DTF Stop Freq" softkey.
► Enter the start frequency and stop frequency you need.
The R&S ZVH adjusts the frequency range according to your settings.
Note that the distance from start to stop frequency has to be equal to the span.
The R&S ZVH adjusts the values if you enter a frequency range that is not the
same as the span.
2.2.2.2 Setting the Start and Stop Distance for DTF Measurements
The start and stop distance define the scaling of the horizontal axis.
In the default setting, the horizontal axis start at 0 m and covers the maximum or stop
distance you have set. By adjusting the start and stop distance, you can zoom in to a
particular section of the cable for further analysis.
► Press the FREQ/DIST key.
► Press the "Start Dist" or "Stop Dist"softkey.
► Enter the distance you want the display to start from and to end at.
In auto span mode, the R&S ZVH adjusts the frequency settings in order to get the
best display resolution.
If you have set the span manually, the R&S ZVH zooms in to the section of the
cable you have defined. The results maintain their accuracy in that case.