![Rohde & Schwarz ZVH-K1 Operating Manual Download Page 82](http://html.mh-extra.com/html/rohde-and-schwarz/zvh-k1/zvh-k1_operating-manual_1477901082.webp)
R&S ZVH
Cable and Antenna Test Mode
Analyzing Measurement Results
Operating Manual 1309.6946.12 - 09
67
2.3.2.6 Using Marker Search Limits
The R&S ZVH allows you to use only a limited section of the trace for the "Set to
Peak", "Set to Next Peak" and "Minimum" functions.
► Press the MKR
key.
► Press the "Search Limits" softkey.
► Select the "Search Limits On/Off" menu item.
► Confirm the selection with the ENTER key.
The R&S ZVH activates the marker search limits.
An [X] indicates an active search limit. Two vertical red lines show the lower and upper
limits in the diagram.
By default, the search limit range is over the whole span.
► Press the "Search Limits" softkey
► Select the "Lower Limit" menu item.
► Confirm the selection with the ENTER key.
The R&S ZVH opens an input field to define the lower limit of the search range.
► Enter the lower limit.
► Confirm the entry with one of the unit keys.
If the span is wide enough, the R&S ZVH displays a red vertical line to indicate the
lower limit.
► Define the upper search limit the same way.
Deactivating marker search limits
► Press the "Search Limits" softkey.
► Select the "Search Limits On/Off" menu item.
► Confirm the selection.
The "Search Limits" softkey turns grey again and in the "Search Limits" menu, the
[X] is no longer displayed.
2.3.3 Using Display and Limit Lines
The display and limit line functionality in CAT mode is the same as that of the spectrum
analyzer mode.
For more information on using the display line see "
For more information on using limit lines see "