![Rohde & Schwarz ZVH-K1 Operating Manual Download Page 119](http://html.mh-extra.com/html/rohde-and-schwarz/zvh-k1/zvh-k1_operating-manual_1477901119.webp)
R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Performing Spectrum Measurements
Operating Manual 1309.6946.12 - 09
104
4.1.8 Measuring the AM Modulation Depth
The AM Modulation Depth measurement analyzes AM modulated signals and
calculates the modulation depth of the signal using the measurement results. Note that
the measurement works properly only if you apply an AM modulated signal.
► Press the MEAS key.
► Press the "Meas Mode" softkey.
The R&S ZVH opens the measurement menu.
► Select the "AM Modulation Depth" menu item.
The R&S ZVH starts to measure the AM modulation depth.
After you have started the measurement, the R&S ZVH places three markers on the
trace. The first marker is placed on the peak power level. The R&S ZVH assumes that
position as the level of the carrier. The second and third markers are delta markers.
These are placed symmetrically on the adjacent peak values to the left and right of the
carrier.
Screen layout for the AM modulation depth measurement
1
Marker list
2
Modulation depth
3
Trace
4
Threshold line
5
Markers (blue lines)
6
AM modulation depth softkey menu
By default, delta marker 2 is active for editing. If you move the delta marker to another
position, the other delta marker will be moved by the same distance relative to the
normal marker. Note that this happens only if you move delta marker 2 (D2). When
moving delta marker 3 (D3), only this marker is repositioned.