R&S ZVH
Network Analyzer Mode (R&S ZVH-K42)
Vector Voltmeter (R&S ZVH-K45)
Operating Manual 1309.6946.12 - 09
201
6.4.3 Evaluating the Results
6.4.3.1 Comparing Results
If you perform measurements on different DUTs whose results you'd like to compare,
you can save the current results as reference values.
► Press the MEAS key.
► Press the "
Ref" softkey.
The R&S ZVH saves the results
that it got prior to pressing the
softkey as the reference result for
future measurements.
The results show the difference of
the current measurement and the
reference measurement. The unit
for reference measurements is
always dB.
The reference measurement results are always displayed below the current
measurement results for better orientation.
Note that reference measurements are only available for the "Vector Voltmeter"
measurement format.