Network Analyzer Mode (R&S ZVH-K42)
Performing Vector Measurements
Operating Manual 1309.6946.12 - 09
185
6.2.2 Measuring the Reflection
The example included here is based on a reflection measurement of the two-port filter
that was also used for the transmission measurement.
You can skip the preset, vector measurement selection and frequency settings if you
have already set up the R&S ZVH for the measurement. You can also skip the
calibration, if you have performed the appropriate calibration already.
Preset the R&S ZVH
Before starting the measurement procedure, preset the R&S ZVH to restore the default
configuration and connect the filter between the measurement ports.
► Press the PRESET key.
► Connect the DUT.
Select the type of reflection measurement
► Press the MEAS key.
► Press the "Reflection S11" or "Reflection S22" softkey for reflection measurements
on port 1 or reflection measurements on port 2.
Simultaneous display of several results
You can also display two results simultaneously in a single diagram. In that case, the
R&S ZVH shows two traces with a different color in the result display.
See “
” for more information.
Define the frequency parameters
Before calibrating the measurement, you should define the frequency parameters to
avoid interpolation of the results.
► Press the FREQ/DIST key.
The R&S ZVH opens an input field to set the center frequency.
► Enter a frequency of 1950 MHz.
► Press the FREQ/DIST key.
► Press the “Span" softkey.
The R&S ZVH opens an input field to define the span.
► Enter a span of 130 MHz to get a high resolution of the results.