Working with the Measurement Wizard
Using the Measurement Wizard
Operating Manual 1309.6946.12 - 09
74
3.2.2 Performing a Sequence of Measurements
Now that you have updated all parameters concerning the measurement task, you can
start the measurement procedure.
► Press the "Start Meas Set" softkey.
If you have not yet calibrated the R&S ZVH for the measurement, it asks you to
perform the calibration routine. The stages of the calibration depend on the defined
calibration routine.
For more information see "
Calibration pool
The R&S ZVH adds the calibration data of each calibration you perform to a calibration
pool available on its internal memory. Each dataset in the calibration pool is unique for
a particular calibration method and a particular measurement configuration.
Note that the calibration is only valid for the instrument it has been performed on.
Before you start a measurement sequence with the measurement wizard, the
R&S ZVH compares the contents of the calibration pool to the measurement
configurations and required calibration methods in the measurement sequence.
● If the R&S ZVH has already been calibrated for a particular configuration, it
restores that data. Another calibration is not necessary.
● If the R&S ZVH has not been calibrated for a particular configuration yet,
calibration becomes necessary. The new calibration data is added to the pool.
Therefore, calibration is necessary only for calibration methods that have not been
used for a particular measurement configuration before.
However, it is recommended to perform calibration on a regular basis to keep
measurements as accurate as possible.
You can delete obsolete calibration data from the folder:
\Public\CalibrationPool
.
Measurement sequence
After having successfully calibrated the R&S ZVH, it starts to go through the
measurements that are part of the measurement set. The sequence of measurements
is as defined with R&S ZVHView.
Before each measurement, the
R&S ZVH shows a message box.
The message box contains the information and instructions on how to prepare and
perform the measurement that you have defined with the R&S ZVHView software.
► Make the necessary preparations, like connecting the cable.
► Press the "Confirm" softkey.