R&S ZVH
Network Analyzer Mode (R&S ZVH-K42)
Evaluating the Results
Operating Manual 1309.6946.12 - 09
195
The reference position is a number between 0 and 10 with 0 being the grid line at the
bottom and 10 being the grid line at the top of the diagram.
► Press the SCALE/AMPT key.
► Press the "Range/Ref Pos" softkey.
► Select the "Ref Position:" menu item.
The R&S ZVH opens an input field.
► Enter the new position of the reference value.
The R&S ZVH moves the trace accordingly and moves the triangle to the new
position.
Changing the display range
The display range is the value range the vertical axis covers. The default display range
depends on the measurement format. For more information see "
When you change the display range, the reference value and position remain the
same. Merely the scale of the vertical axis is compressed or expanded.
► Press the SCALE/AMPT key.
► Press the "Range/Ref Pos" softkey.
► Select the display range you need from the submenu.
The R&S ZVH adjusts the vertical axis accordingly.
Automatic scaling of the vertical axis
All measurement formats except the Smith chart feature an automatic scale function.
If you use automatic scaling, the R&S ZVH automatically sets the display range in a
way that the trace best fits on the display.
► Press the "Auto Scale" softkey.