R&S ZVH
Working with the Measurement Wizard
Using the Measurement Wizard
Operating Manual 1309.6946.12 - 09
75
The R&S ZVH performs the measurement as defined in the dataset and
measurement set. When finished, it shows the measurement results and says
.
Note that it is not possible to change any measurement parameters while using the
measurement wizard. Marker functionality, scaling parameters and (if configured
that way) cable settings are, however, available.
Changing cable characteristics
If necessary, you can define different cable characteristics (cable model and length)
after each individual measurement. This is useful, for example, if you want to do the
same measurements on different cables at one go.
This feature is available when you turn on "Prompt User to Change Cable Settings" in
the "Wizard Set Editor" of the R&S ZVHView software.
When this function is on, the R&S ZVH asks you to select new cable characteristics
during the preparation of a measurement.
► Press the "Yes" softkey if changes are required and select new cable
characteristics before proceeding with the measurement.
► Press the "No" softkey if changes are not required and proceed with the
measurement.
After each measurement step, you have several options:
-
Continue with the next measurement ("Continue" softkey).
Finishes the current measurement and begins with the next measurement by
showing the necessary preparations.
-
Repeat the current measurement ("Repeat Meas" softkey).
Repeats the current measurement, e.g. if the results don't match your
expectations and you want to validate the results.
-
Interrupt the measurement set sequence ("Interrupt Wizard" softkey).
Interrupting the sequence of measurements may become necessary if the
measurement doesn't yield the expected results (e.g. violated limit lines).
In that case, you can interrupt the measurement sequence and try to find the
origin of the problem by using different settings or measurements than those
defined in the wizard.
When you interrupt the wizard sequence, the complete functionality is
available as if you would not use the wizard.
When you interrupt a measurement sequence, the R&S ZVH keeps the results
of measurements you have already performed.
When you are finished reconfiguring the measurement, press the WIZARD key