64
2.3.2.2 Positioning a Delta Marker
When a normal marker is already in use, you can add delta markers.
► Press the MARKER key.
► Press the "New Marker" softkey.
The R&S ZVH activates a delta marker and positions it on the next maximum level
that has been measured. In addition, the delta marker input field opens.
The R&S ZVH adds the delta marker to the marker list and shows the marker
position relative to the normal marker (M1).
You can perform the following actions:
● Enter a delta marker position with the number keys and confirm the entry with one
of the unit keys.
● Change the delta marker position with the rotary knob or the cursor keys.
► Confirm the delta marker position with the ENTER key.
The delta marker input field closes.
► To add more markers, press the "New Marker" softkey several times until you have
the number of markers you want in the display.
2.3.2.3 Selecting the Marker Type
When you add new markers, they will be delta markers by default. Their coordinates
are relative to the first marker (M1). You can turn delta markers into normal markers if
you need absolute information about the marker position.
► Press the MARKER key.
► Select the delta marker you want to convert with the "Select Marker" softkey.
The corresponding marker symbol turns red and the marker input field opens.
► Press the "Marker Type" softkey.
The delta marker turns into a normal marker. Its label changes accordingly (e.g. D2
to M2) and its coordinates are now absolute values.