![Rohde & Schwarz ZVH-K1 Operating Manual Download Page 183](http://html.mh-extra.com/html/rohde-and-schwarz/zvh-k1/zvh-k1_operating-manual_1477901183.webp)
R&S ZVH
Power Meter (R&S ZVH-K9)
Using the Internal Power Meter (R&S ZVH-K19)
Operating Manual 1309.6946.12 - 09
168
5.3 Using the Internal Power Meter (R&S ZVH-K19)
The R&S ZVH also supports power measurements without using a power sensor. In
that case, you can connect the DUT directly to the R&S ZVH and still perform accurate
channel power measurements.
The screen layout is the same as is described in "
Performing and configuring channel power measurements
The configuration of channel power measurements without a power sensor is similar to
measurements with a power meter.
The following features are available:
● Defining the frequency
● Zeroing the measurement
● Selecting the unit
● Defining the reference level
● Taking additional loss or gain into account
For more information see "Performing and Configuring Measurements" on page 160.
Defining the channel bandwidth
In addition, you can select the channel bandwidth.
► Press the MEAS key.
► Press the "Channel BW" softkey.
The R&S ZVH opens an input field.
► Enter the required channel bandwidth.
The R&S ZVH performs a measurement on the selected channel. Note that it is not
possible to change the measurement time, resolution bandwidth and frequency span.