R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Performing Spectrum Measurements
Operating Manual 1309.6946.12 - 09
85
To prevent an overload, perform the measurement with the largest resolution
bandwidth possible using the peak detector. If you set these parameters, it is not
possible for the trace to exceed the reference level.
To simplify operation and to prevent incorrect measurements, the R&S ZVH has an
automatic routine for setting the reference level.
► Press the SCALE/AMPT key.
► Press the "Level Adjust" softkey.
The R&S ZVH performs a measurement to determine the ideal reference level.
It uses a resolution bandwidth of 1 MHz, a video bandwidth of 1 MHz and the peak
detector. After finishing the automatic measurement, the R&S ZVH sets the ideal
reference level.
4.1.3.3 Setting the Channel Bandwidth
The channel bandwidth specifies the frequency range around the center frequency,
over which the R&S ZVH performs the power measurement.
► Press the MEAS key.
► Press the "Chan BW" softkey.
The R&S ZVH opens an input field to specify the channel bandwidth.
► Enter the channel bandwidth you need.
The R&S ZVH sets the appropriate span for the channel bandwidth that you have
entered (span = 5 x channel bandwidth). This ensures that no incorrect channel
power measurements are made.
The minimum channel bandwidth that you can set is 2 kHz at a span of 1 kHz.
4.1.3.4 Defining the Percentage of Occupied Bandwidth
By default, the power percentage inside the occupied bandwidth is 99 %, a number
most standards require. If you measure standards that specify another power
percentage you can change that value.
► Press the MEAS key.
► Press the "% Power BW" softkey.
The R&S ZVH opens an input field to define the power percentage.
► Enter the power percentage you need.
The value you enter here is the power percentage that has to be inside the channel
bandwidth relative to the power over the total span (the percentage of the total
power).
The R&S ZVH now displays the occupied bandwidth graphically in the trace
window and numerically above the trace window.