R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Performing Spectrum Measurements
Operating Manual 1309.6946.12 - 09
80
4.1.2.1 Selecting the Standard
If you need to perform measurements that are conform to a telecommunications
standard, you can activate one of the predefined standards that are already stored in
the R&S ZVH memory. However, you can create new configurations to perform
measurements on other standards as well.
► Press the MEAS key.
► Press the "Standard" softkey.
The R&S ZVH opens a dialog box to select the standard.
► Select one of the available standards.
► Confirm the selection with the "Select" softkey.
The R&S ZVH loads the configuration of the selected standard. It automatically
sets the optimal span, resolution bandwidth, video bandwidth, sweep time and
detector for the standard.
If the measurement is not according to the selected standard anymore (e.g. if you
make changes to a parameter), the R&S ZVH puts a red dot in front of the
standard display (
).
You can create and edit standards with the R&S ZVHView software and transfer them
to the R&S ZVH via the USB or the LAN interface. The number of standards the
R&S ZVH can store in its memory depends on the number of other data sets stored on
the R&S ZVH. For more information see "
4.1.2.2 Setting the Reference Level
The reference level is the power level that the R&S ZVH expects at the RF input. When
selecting the reference level, make sure that you do not overload the R&S ZVH by
applying a signal whose power exeeds the maximum reference level.
As the power is measured with a small resolution bandwidth compared to the signal
bandwidth, it is still possible to overload the R&S ZVH, even if the trace is within the
measurement diagram. To prevent an overload, perform the measurement with the
largest resolution bandwidth possible using the peak detector. If you set these
parameters, it is not possible for the trace to exceed the reference level.
To simplify operation and to prevent incorrect measurements, the R&S ZVH has an
automatic routine for setting the reference level.
► Press the SCALE/AMPT key.
► Press the "Level Adjust" softkey.
The R&S ZVH performs a
measurement to determine the
optimal reference level.
It uses a resolution bandwidth of 1 MHz, a video bandwidth of 1 MHz and the peak
detector. After finishing the automatic measurement, the R&S ZVH sets the optimal
reference level.