R&S ZVH
Spectrum Analyzer Mode (R&S ZVH-K1)
Configuring Spectrum Measurements
Operating Manual 1309.6946.12 - 09
135
4.2.4.2 Selecting the Sweep Mode
The sweep mode is the way the R&S ZVH performs the measurement.
In its default state, the R&S ZVH measures continuously. In this mode, the R&S ZVH
automatically repeats the sweep in the defined range of the horizontal axis (frequency
or time) and updates the trace accordingly after it has finished with one sweep.
In some cases it may be sufficient to get the results over a single sweep only, e.g. if a
particular trigger condition is met. In single sweep mode, the R&S ZVH performs the
sweep a certain number of times (depending on the number of averages you have set)
over the defined range of the horizontal axis (frequency or time) and then stops
measuring. It performs another sweep only after you tell it to. For more information on
setting the number of sweeps included in a single sweep see "
Mode (Average)
".
► Press the SWEEP/BW key.
► Press the "Trigger" softkey.
► Select the "Single Sweep menu item to perform measurements in single sweep
mode or the "Continuous Sweep" menu item to perform measurement
continuously.
4.2.4.3 Working with Trigger Functionality
If you have to perform measurements according to certain signal conditions, you can
use a trigger. A trigger responds to certain events. If a trigger is active, the R&S ZVH
starts to measure if the trigger conditions are met. The trigger can be generated either
externally or internally.
Selecting the trigger source
► Press the SWEEP/BW key.
► Press the "Trigger" softkey.
The R&S ZVH opens a submenu to select the trigger source.
► Select the trigger source you need.
The R&S ZVH activates the trigger.
The R&S ZVH provides the following trigger functions.
●
Free Run
A new sweep starts on completion of the previous sweep. This is the default state
of the R&S ZVH.
●
Video Trigger
A sweep starts when the video voltage exceeds a particular level. The video trigger
is available only in the time domain (span = 0).